Ventilator Control and Drive
Theory of Operation
1 - 28
1.5.3
Battery
The sealed lead-acid battery is maintenance free, and has a maximum recharge time of 8
hours when fully discharged. The run time is at least 45 minutes with a fully charged battery.
To prevent unintended loss of battery operation, the recommended replacement period is
every 3 years.
1.5.4
Power Supply
The power supply provides power to the machine and relevant controls.
8.
APE
Auxiliary Power Enable
9.
CHAR_SIGNAL
Charge Status, No connection
10.
/MUTE
Mute, No connection
Power Switch, J5
PIN NAME FUNCTION
1. GND Ground
2. NC
No
Connection
3.
PWON
Power On Enable, Low active
Test Point definitions
DESIGNATOR NAME
FUNCTION
T1
24V
24V DC Input
T2 PWBUS Power
Bus
T3 A12V Analogue
12V
T4 P12V Power
12V
T5
+5V
Logic Power 5V
T6 5V/4.8V
Power
5V/4.8V
T7 BAT+ Battery
Input
+
T8
30V
30V for charger
T9 PWAUX Power
Auxiliary
T10 CHARGE Charge
Status
T11 5VAUX 5V
Auxiliary
T12 VRs
Charger
Feedback
Led Designation
DESIGNATOR LED
COLOR
STATUS
FUNCTION
DS1
GREEN
24V DC available
DS2
GREEN
Power Bus On
DS3 GREEN Charge
Power Control, J2 (Continued)
PIN NAME FUNCTION
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...