1 - 17
Theory of Operation
Ventilator Control and Drive
23. AGND
Analog
Ground
24. AGND
Analog
Ground
25. AGND
Analog
Ground
Bus to BDU, J96
PIN NAME
FUNCTION
1 - 3
5V
5V Power
4 - 6
GND
Ground
Test Point Definition
DESIGNATOR
NAME
FUNCTION
RANGE
T1 FIO
2
ADC input of O
2
Sensor
0.25-0.51V (In AIR)
T2
Pgas
Pressure switch of gas supply,
low active
T3
FL_INSP
Inspire Flow signal
0.20-0.30V (0 flow)
T4
Pair
Absolute Pressure of air way
1.7-2.1V (atmospheric
pressure)
T5
FL_EXP
Expire flow signal
0.20-0.30V (0 flow)
T6
Fm_fl
Flow of flowmeter
0.25-0.51V (flowmeter off)
T7
AD-Exhale
PEEP valve, current feedback
Level depends on PEEP
setting and vent mode. No
activity may indicate open
connection from driver to
valve (ie. bad cable)
T8
Volt
Internal power inspect
T9
MUTE
Not used
T10
M-EN
Manual/Auto mode select
T11
V-EN
Valves enable Input
Global valve enable signal.
Must be active for valves to
operate.
T12
EX-DA
Exhale DA output
T13
In-DA
Inhale DA output
T14
GS_ST
gas supply status, high active
T15
XVI
unamplified version of T7
T16
IVI
unamplified version of T28
T17
MVI
Manual-Valve current feedback
T18
IVG
Proportional valve drive signal,
drives T16 and T28
Level depends on machine
settings.
T19
XVG
PEEP valve drive signal - drives
T15 and T7
Level depends on PEEP
setting and vend mode.
T20
PADJ
Power 4.8V
4.8V
T21
P24V
Power 24V
24V
T22
P12V
Power 12V
12V
Signals and Power to Sensor Board, J30 (Continued)
PIN NAME
FUNCTION
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...