Tr
oubleshooti
ng Char
t
Repair
Inf
or
m
ation
3 - 10
0070-10-0683
AS3000™ Ser
High FiO
2
Runtime
1
Oxygen concentration is higher than the
upper limit set for the Oxygen Concentration
alarm.
2
Oxygen calibration is inaccurate.
1
Adjust the upper limit set for the Oxygen
Concentration alarm or the actual oxygen
concentration.
2
Calibrate the oxygen concentration sensor
again.
APNEA
Runtime
1
Breathing hoses are disconnected.
1
Check the breathing hoses.
Low Battery
Runtime
1
Battery capacity is low.
1
Connect the unit to AC power.
High Minute Volume
Runtime
1
Minute ventilation is higher than the upper
limit set for the Oxygen Concentration
alarm.
1
Adjust the upper limit set for the Minute
Ventilation alarm.
Low Minute Volume
Runtime
1
Minute ventilation is lower than the lower
limit set for the Minute Volume alarm.
1
Adjust the lower limit set for the Minute Volume
alarm.
APNEA Backup
Runtime
1
PS mode is not triggered and APNEA
backup ventilation is given according to the
minimum breathing frequency that has been
set.
1
No action needed.
High PEEP
Runtime
1
The expiration time is too short or expiration
limb is occluded
2
The expiration valve (PEEP) is defective.
1
Check the time setting and expiration pipeline.
2
Check the expiration valve (PEEP).
Keyboard communication
Failure
Startup Test
Runtime
1
The signal wire is disconnected or
damaged.
2
The keyboard board program crashed.
1
Check the signal wire.
2
Power on again.
AC Power Failure
All
1
AC power is not connected or a fuse is
tripped.
1
Connect AC and check the fuses.
O
2
Cal Due
Startup Test
Runtime
1
The oxygen sensor has not been calibrated
for more than 3 days.
1
Calibrate the oxygen sensor.
Tidal volume displayed by
the
AS3000
is smaller than
the value displayed on
bellow.
Runtime
1
This caused by the compliance and
compensation of resisting force between the
airway and patient.
1
No action needed.
Bellows is deflated and at the
bottom of its travel during
runtime, after initial power-up
and Leak Test.
Runtime
1
This is caused by a disconnected breathing
circuit, a removed absorber canister, a
removed O
2
sensor or sensor adapter,
check valve dome rings, or no fresh gas
flow while ventilator is running.
1
Install a breathing circuit and connect the Y-
piece to the Test port. Select CMV mode and
confirm the selection. Immediately push the O
2
Flush button until the bellows is completely
filled. Return to STANDBY mode and remove Y-
piece from Test port.
MESSAGE/ISSUE
OCCURRENCE
CAUSE
SOLUTION
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...