Check Valve Cleaning
Periodic Maintenance
6 - 6
2. Remove Absorber Guide Rods and Mounting Plate:
1.
Loosen the locking nuts on (2) Absorber Guide Rods.
2.
Remove the (2) Absorber Guide Rods by unscrewing them from the Breathing System
Block.
3.
Remove Gasket and Screen from Absorber Mounting Plate.
4.
Remove (3) philips head screws with (3) associated washers securing the Absorber
Mounting Plate.
5.
Remove the Absorber Mounting Plate
FIGURE 6-3
Absorber guide rods and mounting plate
Absorber Guide
Rod
Locking Nuts
Gasket and
Screen