6 - 39
Periodic Maintenance
Cleaning
6.9
Cleaning
6.9.1
Cleaning and Disinfecting the AS3000
Before cleaning, switch off the
and disconnect it from the mains.
NOTE:
See the “AS3000 Operating Instructions”
(P/N: 0070-00-0684-XX) for more in-depth cleaning
instructions.
6.9.2
Cleaning and Sterilizing the Breathing System and Components
6.10
Preoperative Checklist
This checklist should be conducted before administering anesthesia.
NOTE:
This is a guideline which can be modified to accommodate
variations in local clinical practice. Such local modifications
should have appropriate peer review.
If an anesthetist uses the same machine in successive cases, this checkout need not be
repeated or may be abbreviated after the initial checkout.
1. Inspect the system for:
a.
Identification number
b.
Valid inspection sticker
c.
Damage to flowmeters, vaporizers, gauges, and supply hoses
d.
Complete Breathing System with adequate CO
2
absorbent Pre-Paks or loose fill
e.
Correct mounting of gas cylinders in the yokes
f.
Presence of the Tank Wrench
Cleaning method and chemical
COMPONENT
SOAP
WATER
CHEMICALS
CIDEX
STERILIZATION
STEAM
AUTOCLAVE
MAXIMUM
TEMPERATURE
(°F/°C)
Absorber canister
yes
yes
no
Airway pressure gauge
no
no
no
APL Valve
yes
yes
no
Valve cover and valve nut
yes
yes
no
Breathing System module
(without the above
mentioned)
yes
yes
yes
273°F/134°C
AGSS yes
yes
no
The components in the above table should be rinsed in warm water thoroughly and dried in air after cleaning
or sterilizing.
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...