2 - 23
Installation Guide
Installation Checkout Procedure
23. Verify that the drive gas pressure loss alarm signals activate and N
2
O
stops and AIR continues to flow.
a.
Set the O
2
flow for 1 L/min.
b.
Set the N
2
O flow to 1 L/min.
c.
Set the AIR flow to 1 L/min.
d.
Disconnect the O
2
line supply to the
, and Close the O
2
cylinder.
e.
Verify that the flow of N
2
O stops when O
2
is lost.
f.
Verify that AIR continues to flow.
g.
Verify the “O
2
Supply Failure” message appears on screen, a steady audio tone
should sound.
h.
Reconnect O
2
line pressure source to the
.
i.
Verify that the alarms are reset.
j.
Verify N
2
O flow has resumed once O
2
pressure returns.
24. Verify that the battery operation and charging icons appears.
a.
Set the vent mode to Standby.
b.
Disconnect AC line cord.
c.
Verify that the battery icon appears on the screen and indicates the charge level of
the battery.
d.
Verify that the line cord icon is flashing.
e.
Verify that the flow meter backlight is illuminated during battery operation.
f.
When the battery icon is in a not full state, reconnect the AC line cord.
g.
Verify that the battery charge icon appears on the screen.
h.
Verify that the flow meter backlight is illuminated during AC voltage operation.
25. Verify operation of the work light.
a.
Turn on the work light located on the bottom side of the top panel.
b.
Verify that it lights in both on positions.
26. Verify that the Auxiliary O
2
and AIR Flowmeters operates.
a.
Verify an AIR flow of 15 L/min can be obtained by connecting the auxiliary AIR hose
to the pressure source and opening the flow meter.
b.
Verify an O
2
flow of 15 L/min can be obtained by connecting the auxiliary O
2
hose
to the pressure source and opening the flow meter.
27. Verify that the Suction Regulator operates in LINE and REGULATE
positions.
a.
Set the suction regulator's selection dial to
LINE
.
b.
Verify maximum suction vacuum on the regulator's gauge.
c.
Set the suction regulator's selection dial to
REGULATE
.
d.
Verify the suction vacuum is adjustable on the regulator's gauge.
e.
Set the suction regulator's selection dial to
OFF
.
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...