Installation Checkout Procedure
Installation Guide
2 - 24
28. Verify that the Breathing System heats to body temperature.
a.
Operate the
on AC operation for approximately 40 minutes.
b.
Verify the Breathing System has heated up to body temperature
29. Complete electrical safety inspection.
NOTE:
Perform the electrical safety inspection as the last step after
completing a repair or after routine maintenance. Perform
this inspection with all covers, panels, and screws installed.
a.
Withdraw the Power cable(s) from the convenience receptacles at the rear of the
.
b.
Plug the
into a Safety Analyzer.
c.
Connect the case ground lead of the analyzer to the U-blade ground of one of the
convenience receptacles. Perform the following tests with the case grounded:
• Normal polarity
• Normal polarity with open neutral
d.
Perform the following tests with the case ungrounded:
• Normal polarity
• Normal polarity with open neutral
• Reverse polarity
e.
Verify that the maximum leakage current does not exceed 300 µA (0.3 mA).
NOTE:
Ground Resistance (between the U-blade ground on any
convenience outlet to the U-blade ground on the AC line
cord).
f.
Plug the
into the safety analyzer.
g.
Attach the resistance-measuring probe on the analyzer to the
U-blade
ground on any convenience outlet.
h.
Invoke the resistance function on the safety analyzer, following the instructions for the
analyzer.
i.
Verify that resistance to ground is less than 0.2 Ohms (200 mOhms).
30. Check the vaporizer interlock.
a.
Attach two vaporizers to the Vaporizer Mounting Manifold and lock them in place.
b.
Rotate either of the vaporizer’s dial to
3%
agent.
c.
Verify that the other vaporizer dial cannot be rotated to a setting.
d.
Set both vaporizer’s dials to
0
.
e.
Rotate the other vaporizer dial to
3%
.
f.
Verify that the first vaporizer dial cannot be rotated.
g.
Rotate both vaporizer dials to
T
and remove both vaporizers.
h.
Verify that the locking spring is intact.
i.
Reconnect both vaporizers to the Vaporizer Mounting Manifold
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...