AS3000™ Ser
“Inspiration Sensor . . . FAIL”
or
“Inspiration Sensor . . . FAIL
Try re-starting the system with
all gas flows OFF” (UI
version 2.25 and higher)
Startup test
1
Fresh gas is flowing during the startup
procedure.
2
The tubing kinked or occluded.
3
The sensor board is defective.
1
Verify that no fresh gas is flowing during the
startup procedure.
2
Check the tubing for kinks and remove any
occlusions.
3
Replace the sensor board and then perform
each item of the calibration and startup test
again.
“O
2
Supply . . . FAIL”
Startup test
Runtime
1
The gas source is not connected.
2
The gas supply pressure switch or its cable
is damaged.
3
The amplifier board is defective.
1
Check gas source.
2
Check and replace the switch or cable of the
gas source.
3
Replace the amplifier board, and then perform
each item of the calibration and startup test
again.
“AC Power Failure”
Startup test
Runtime
1
AC power is not connected.
2
A fuse is tripped
1
Check the connection of AC power.
2
Replace the fuse.
“Low Battery”
Runtime
1
Battery capacity is low.
2
The battery is defective.
3
The connection wire or power board is
defective.
1
Charge the battery by connecting AC power.
2
Replace the battery.
3
Check the connection wire or replace the
power board.
“Alarm Speaker . . . FAIL”
Startup test
1
The connection wire to the speaker is
disconnected or damaged.
2
The speaker is damaged.
3
The keyboard board is defective.
1
Check whether the connection wire of the
speaker is disconnected or damaged.
2
Replace the user interface assembly.
3
Replace the user interface assembly.
“Software Mismatch”
or
“Software Mismatch or
Failure to Shut Down
Completely” (UI version 2.25
and higher)
Startup test
1
Power is recycled to the unit too quickly, not
allowing the memory to fully clear.
1
Shut down the unit and wait until the backlight
is fully extinguished before restoring power.
Charging state is always
displayed.
Runtime
1
Battery capacity is low.
2
The connection wire to the battery is
disconnected or damaged.
3
The battery is defective.
1
Charge the battery by connecting AC power.
2
Check whether the connection wire to the
battery is disconnected or damaged.
3
Replace the battery.
During startup leak test, the
bellows moves when it
shouldn’t
Startup test
1
A solenoid failure or wiring error.
1
Correct the wiring or replace the solenoid
valve.
In leakage detection, The APL
Valve detection pressure
displayed on the pressure
gauge does not stay within
28 - 32 cmH
2
O.
Startup test
1
The Breathing System pneumatic hose leaks
gas when in manual mode.
2
Pressure gauge failure.
3
The APL Valve is out of spec
1
Replace the Breathing System pneumatic hose
2
Replace the pressure gauge
3
Replace the APL Valve
MESSAGE/ISSUE
OCCURRENCE
CAUSE
SOLUTION
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...