Calibration Procedures
Calibration
5 - 6
5.5.2
Proportional Valve Regulator Calibration
1.
Open the gas cover panel.
2.
Remove the drive pressure source from the
3.
Disconnect the input hose from the Proportional Valve.
4.
Connect a Digital Pressure Meter to the proportional valve input hose.
5.
Reconnect the drive pressure source to the
6.
Set the pressure on the pressure regulator to between 25.1 and 25.7 psi by pulling
down on the pressure regulator knob and rotating it until the desired pressure is
reached. (see FIGURE 5-1)
7.
Push the regulator knob back up once the desired pressure is reached to lock it in place.
8.
Remove the drive pressure source from the
9.
Remove the Pressure Meter and reconnect the input hose to the Proportional Valve.
10.
Reconnect the drive pressure source to the
FIGURE 5-1
The Proportional Valve Regulator
Proportional valve input hose