MS to MSMS Switching
Switch Criteria
MSMS scanning
commences:
•
If
TIC
is
selected,
and the TIC
of the
spectrum
rises above
the specified
Threshold
.
•
If
Intensity
is selected,
and the
intensity of
the largest
peak rises
above the
specified
Threshold
.
When a peak top
is found, no other
peaks are looked
for within the
specified
Detection Window
.
Currently
Number of Components
is set to 1 and can not be changed. The
number of non coeluting precursors in a single run is not limited.
Precursor Selection
If
Automatic
is selected all valid masses satisfying selection criteria are monitored.
If
Include Masses Only
is selected only masses in the include list (see below) are
monitored.
If
Include Masses and Automatic
is selected masses on the include list are given
priority. If no precursors are found then other valid masses are monitored.
A mass is valid if it is not on the exclude list (see below), and it satisfies the
precursor selection criteria.
Data Acquisition
Page 95
Quattro Ultima
User's Guide
Summary of Contents for Quattro Ultima
Page 2: ...Quattro Ultima User s Guide...
Page 68: ......
Page 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Page 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Page 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Page 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Page 222: ...Reference Information Page 222 Quattro Ultima User s Guide...