MS Operating Modes
MS1
Collision Cell
MS2
MS
Resolving
RF Only (Pass all masses)
MS2
RF Only (Pass all masses)
Resolving
The MS1 mode, in which MS1 is used as the mass filter, is the most common and
most sensitive method of performing MS analysis. This is directly analogous to using
a single quadrupole mass spectrometer.
The MS2 mode of operation is used, with collision gas present, when switching
rapidly between MS and MS-MS operation. It also provides a useful tool for
instrument tuning and calibration prior to MS-MS analysis, and for fault diagnosis.
MS-MS Operating Modes
The basic features of the four common MS-MS scan functions are summarised below.
MS1
Collision
Cell
MS2
Daughter Ion
Spectrum
Static
(parent mass selection)
RF only
(pass all
masses)
Scanning
Parent Ion
Spectrum
Scanning
Static
(daughter mass
selection)
Multiple Reaction
Monitoring
Static
(parent mass selection)
Static
(daughter mass
selection)
Constant Neutral
Loss Spectrum
Scanning (synchronised
with MS2)
Scanning (synchronised
with MS1)
Instrument Description
Page 18
Quattro Ultima
User's Guide
Source
MS1
Collision Cell
MS2
Detector
Summary of Contents for Quattro Ultima
Page 2: ...Quattro Ultima User s Guide...
Page 68: ......
Page 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Page 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Page 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Page 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Page 222: ...Reference Information Page 222 Quattro Ultima User s Guide...