If the correct peaks have been matched then the calibration report can be printed out:
Select
,
from the report display.
If insufficient peaks have been matched or the wrong peaks have been matched see
Calibration Failure
later in this chapter. To accept the calibration:
Select
OK
from the calibration report.
The same procedure is used for the scan speed compensation except that
Scan Speed Compensation
is selected in the dialog box, and the fast scanning file
is used. Note that for the scan speed compensation the default file is
FASTMS1
. If an
alternative filename has been used then this must be selected using the data browser.
Once all three calibrations (static, scanning and scan speed compensation) have been
completed then the instrument can be used for any mass range within the limits of the
scanning calibrations and at any scan speed from 100 to 1000 amu/sec.
Calibrating MS2
The calibration of MS2 is carried out in exactly the same manner as above, except that
data is acquired in MS2 mode instead of MS1.
Using the Instrument
Once all six calibrations (static, scanning and scan speed compensation, each for both
MS1 and MS2) have been completed then the instrument can be used for any mass
range within the limits of the scanning calibrations and at any scan speed from 100 to
1000 amu/sec.
Mass Calibration
Page 134
Quattro Ultima
User's Guide
Summary of Contents for Quattro Ultima
Page 2: ...Quattro Ultima User s Guide...
Page 68: ......
Page 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Page 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Page 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Page 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Page 222: ...Reference Information Page 222 Quattro Ultima User s Guide...