AutoTune
MassLynx can automatically
tune the mass spectrometer in
both APcI and electrospray
ionisation modes. AutoTune
ramps the settings for the
tuning parameters until they
are optimised to give the best
intensity, resolution and peak
shape.
To run AutoTune:
Press
on the tune
page to turn on the API
gas, and select
Operate
.
Choose
AutoTune
from the tune
page
Options
menu.
Press
Setup
to define the
AutoTune setup parameters.
There are two levels of AutoTune:
•
A full AutoTune starts from a
default set of tuning parameters.
•
A maintenance AutoTune starts from the current tuning parameters set in the
tune page and can be quicker than a full AutoTune.
A maintenance AutoTune can only be performed if the instrument is already
reasonably well tuned. If the current tuning is too poor AutoTune gives an error
and request a full AutoTune.
The
Tune Mass
parameter sets the mass to be tuned on. When satisfied with the
AutoTune setup parameters:
Press
OK
to exit.
Press
Start
.
The AutoTune status bar is updated to show the progress of AutoTune.
The following steps are performed:
•
Parameter initialisation and instrument checks
Ensuring that essential status indicators read correctly.
Tuning
Page 62
Quattro Ultima
User's Guide
Summary of Contents for Quattro Ultima
Page 2: ...Quattro Ultima User s Guide...
Page 68: ......
Page 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Page 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Page 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Page 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Page 222: ...Reference Information Page 222 Quattro Ultima User s Guide...