Water
Water is used to cool the turbomolecular pumps.
Nitrogen Gas In
The nitrogen supply (100 psi, 7 bar) should be connected to the
Nitrogen Gas In
push-in connector using 6mm PTFE tubing. If necessary this tubing can be connected
to ¼ inch tubing using standard ¼ inch fittings.
Caution:
Use only PTFE tubing or clean metal tubing to connect between the
nitrogen supply and the instrument. The use of other types of plastic tubing
results in chemical contamination of the source.
Instrument Description
Page 28
Quattro Ultima
User's Guide
ESD
EARTH
FACILITY
OUTLET
INLET
10 AMP (T)
10 AMP (T)
NITROGEN GAS
WATER
CID
GAS
ROTARY PUMP
CONTROL
ROTARY PUMP
CONTROL
OUT
IN
1
0
Mains
Switch
Electrostatic
Discharge
Earth (Ground)
Point
Power
Cord
Fuses
Source
Pumping Line
Backing
Line
To Rotary Pumps
Hexapole
Pumping Line
Rotary Pump
Control
Nitrogen
Gas In
CID Gas
Water
Connections
Exhaust
Gas