![Micromass Quattro Ultima User Manual Download Page 20](http://html1.mh-extra.com/html/micromass/quattro-ultima/quattro-ultima_user-manual_1787371020.webp)
The Parent Ion Spectrum
Typical application:
•
Structural elucidation.
Complementary or confirmatory information (for daughter scan data).
Example:
Parents of the specific daughter ion at
m
195 from reserpine in electrospray
positive ion mode.
The result:
Instrument Description
Page 20
Quattro Ultima
User's Guide
MS1
scanning from
m/z 50 to 650
MS2
static at m/z 195
(daughter mass)
Collision Cell
RF only
(pass all masses)