Caution:
Failure to gas ballast the rotary pump frequently leads to shortened oil
lifetime which in turn may shorten rotary pump lifetime.
Caution:
The instrument should not be vented with the E2M28 manual gas
valve open. The E1M18 manual gas ballast valve should remain open at all
times.
Oil Mist Filter
The rotary pumps are fitted with an Edwards EMF20 oil mist filter which traps oil
vapour from the rotary pump exhaust. The trapped oil is then returned to the rotary
pump during routine gas ballasting. The oil mist filter contains two elements which
require the following maintenance:
•
Change the odour element monthly or whenever the pump emits an oily odour.
•
Change the mist element every time the rotary pump oil is changed.
To change the elements follow the instructions in the Edwards manual.
Foreline Trap
The foreline trap stops oil vapour migrating from the rotary pump to the mass
spectrometer. During normal use, the activated alumina (sorbent) absorbs any oil
vapour, becoming brown in colour. The sorbent should be replaced when this
discolouration reaches the region of the trap furthest from the pump (the vacuum side).
The manufacturers recommend that the sorbent is replaced routinely at three-monthly
intervals.
With the instrument vented and the pump switched off, replace the sorbent as
described in the manufacturer’s literature.
Maintenance and Fault Finding
Page 181
Quattro Ultima
User's Guide
Summary of Contents for Quattro Ultima
Page 2: ...Quattro Ultima User s Guide...
Page 68: ......
Page 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Page 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Page 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Page 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Page 222: ...Reference Information Page 222 Quattro Ultima User s Guide...