![Micromass Quattro Ultima User Manual Download Page 115](http://html1.mh-extra.com/html/micromass/quattro-ultima/quattro-ultima_user-manual_1787371115.webp)
Mass Calibration
Page 115
Quattro Ultima
User's Guide
___________________________________________________________________________________
___________________________________________________________________________________
500
1000
1500
2000
2500
3000
3500
M/z
-0.15
0.65
amu
MS1 Static Calibration
28 matches of 28 tested references. SD = 0.0465
500
1000
1500
2000
2500
3000
3500
M/z
-0.19
0.35
amu
MS1 Scanning Calibration
27 matches of 28 tested references. SD = 0.0459
500
1000
1500
2000
2500
3000
3500
M/z
0.42
1.10
amu
MS1 Scan Speed Compensation Calibration
26 matches of 28 tested references. SD = 0.0538
500
1000
1500
2000
2500
3000
3500
M/z
0.01
1.09
amu
MS2 Static Calibration
28 matches of 28 tested references. SD = 0.0606
500
1000
1500
2000
2500
3000
3500
M/z
-0.17
0.74
amu
MS2 Scanning Calibration
28 matches of 28 tested references. SD = 0.0832
500
1000
1500
2000
2500
3000
3500
M/z
0.00
1.15
amu
28 matches of 28 tested references. SD = 0.0748
MS2 Scan Speed Calibration
Summary of Contents for Quattro Ultima
Page 2: ...Quattro Ultima User s Guide...
Page 68: ......
Page 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Page 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Page 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Page 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Page 222: ...Reference Information Page 222 Quattro Ultima User s Guide...