Dry the cone and nozzle using nitrogen.
If material has built up on the exhaust liner and cleanable baffle:
Remove the cleanable baffle and the exhaust liner.
Clean these components, or obtain replacements.
Fit the cleaned (or the replacement) exhaust liner and cleanable baffle to the ion
block.
Refitting the sample cone and cone gas nozzle is a reversal of the removal procedure.
The source isolation valve is fully open when the screwdriver slot is parallel to
the direction of gas flow. Note that the valve can be rotated past the fully open
position.
Maintenance and Fault Finding
Page 187
Quattro Ultima
User's Guide
Sample
Cone
Cone Gas
Nozzle
Gasket
Exhaust
Liner
Extraction
Tool
Cleanable
Baffle