Exhausts
The exhaust from the rotary pump should be vented to atmosphere outside the
laboratory.
The gas exhaust, which also contains solvent vapours, should be vented via a separate
fume hood, industrial vent or cold trap.
The gas exhaust should be connected using 10mm plastic tubing connected to
the push-in fitting.
Caution:
Do not connect these two exhaust lines together as, in the event of an
instrument failure, rotary pump exhaust could be admitted into the source
chamber producing severe contamination.
CID Gas
Argon is required as collision gas. See
Hardware Specifications
for details.
Power Cord
The mains power cord should be wired to a suitable mains outlet using a standard
plug. For plugs with an integral fuse, the fuse should be rated at 13 amps.
Mains Switch
The mains switch switches mains power to the instrument.
Fuses
Refer to
Maintenance and Fault Finding
for details of rear panel fuses, and all other
instrument fuses.
Rotary Control
Mains power to the two rotary pumps is controlled by the data system using one of
these two sockets. The other socket is connected to the solenoid valve situated in the
oil return tube on the E1M18 pump.
ESD Earth Facility
A suitable wrist band should be connected to this point when handling sensitive
electronic components, to prevent damage by electrostatic discharge.
Instrument Description
Page 29
Quattro Ultima
User's Guide
Summary of Contents for Quattro Ultima
Page 2: ...Quattro Ultima User s Guide...
Page 68: ......
Page 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Page 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Page 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Page 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Page 222: ...Reference Information Page 222 Quattro Ultima User s Guide...