Using a GVF/16 ferrule and the adapter nut, connect the sample capillary to the
LC union, ensuring that both the liner tube and sample capillary are fully butted
into the LC union.
Disconnect the LC connection and feed the sample capillary through the probe,
ensuring that a 0.3mm graphitised vespel ferrule (GVF/003) is fitted.
Using a Rheodyne spanner, gently tighten the adapter nut onto the probe.
Replace the probe tip and adjust so that 0.5mm of sample capillary protrudes
from the probe tip.
Replace the probe end cover and tighten the grub screw to clamp the LC union.
Maintenance and Fault Finding
Page 199
Quattro Ultima
User's Guide