When real-time update is on the display is continually updated with spectra from the
current acquisition. The actual information displayed is determined by selecting one of
the following radio buttons.
•
Latest scan
displays the last acquired scan. This is the default option.
•
Average all scans
updates the display with spectra formed by averaging all
the spectra that have so far been acquired.
•
Average latest scans
updates the display with spectra formed by averaging
the last
n
scans acquired, where
n
is specified in the associated edit control.
Instrument Data Thresholds
MassLynx has several parameters that allow control over how the system
pre-processes data before it is sent to the host computer. These parameters are
contained in the instrument data thresholding dialog.
Instrument data thresholding allows the user to specify the type of data to acquire and
write to disk, and the type of data to discard and not write to disk. Limiting the
amount of data stored on disk can be particularly desirable when acquiring continuum
data and doing long LC runs.
To change data thresholding:
Choose
Set Instrument Threshold
from the tune page
Options
menu.
Make the required changes to the information.
Press
OK
.
These new parameters are downloaded at the start of the next acquisition scan.
Data Acquisition
Page 75
Quattro Ultima
User's Guide
Summary of Contents for Quattro Ultima
Page 2: ...Quattro Ultima User s Guide...
Page 68: ......
Page 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Page 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Page 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Page 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Page 222: ...Reference Information Page 222 Quattro Ultima User s Guide...