Scan Time
specifies the duration of each scan in seconds.
Inter Scan Time
specifies the time in seconds between a scan finishing and the next
one starting. During this period no data are stored.
Pressing
Origin
allows additional information about the sample to be analysed to be
entered into the following fields:
•
Submitter
•
Job
•
Task
•
Conditions
Multiple Samples
The MassLynx top level screen contains a sample list editor for defining multiple
samples which may be used together to perform a quantitative analysis. The list of
samples is set up using a spreadsheet style editor, which can be tailored to suit
different requirements.
To start a multi-sample acquisition:
Set up a sample list (see
MassLynx NT User Guide
,
Sample Lists
for details).
Choose
Start
from the top level
Run
menu, or press
.
This displays the start sample list run dialog.
Check the
Acquire Sample Data
,
Auto Process Samples
and
Auto Quantify Samples
boxes as required.
Enter values in the
Run From Sample
and
To Sample
boxes.
The default is all samples in the list.
Check the
Priority
and/or
Night Time Process
boxes as required.
See the “Getting Started” chapter of the MassLynx manual for details.
Press
OK
.
Repeat the above procedure as required.
Sample lists are added to a queue and run sequentially unless
Priority
or
Night Time Process
has been checked.
The sample which is currently being acquired has a
next to it in the sample
list.
Data Acquisition
Page 71
Quattro Ultima
User's Guide
Summary of Contents for Quattro Ultima
Page 2: ...Quattro Ultima User s Guide...
Page 68: ......
Page 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Page 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Page 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Page 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Page 222: ...Reference Information Page 222 Quattro Ultima User s Guide...