The recommended scan speed for the scan speed compensation is 1000 amu/sec. This
is the maximum scan speed permissible when using thresholded continuum data.
Although continuum is recommended centroided data may be used. It is possible
to scan more quickly in centroided mode, but it is unlikely that a faster
acquisition rate would be needed for general use.
Set
Scan From
to 80 amu and
Scan To
to 1000 amu.
Set
Scan Time
to 0.92 sec and
Inter Scan Delay
to 0.1 sec.
Select
Continuum
as the
Data Type
.
Set
Run Duration
to 2.0 minutes.
Choose any filename for the data.
The filename FASTMS1, the name used during an automatic calibration, is
valid.
Start the acquisition and inject the reference solution.
Manual Calibration
Find the start and end scans of the reference data for each file in the same way
as for the static calibration file.
From the Calibration dialog box select
Process
,
Calibration From File...
.
Select
Scanning
calibration type and
MS1
.
In the lower area the data filename
SCNMS1
should be selected automatically.
If this is not the case, or if an alternative filename has been used for the slow
scanning acquisition, then the correct file can be selected by clicking on
Browse...
.
Enter the start and end scans of the reference data in the
From
and
To
boxes.
Select
OK
to perform the calibration and display the calibration report on the
screen in a similar way to the static calibration.
Compare the acquired and reference spectra to ensure that the correct peaks have
been matched.
Mass Calibration
Page 133
Quattro Ultima
User's Guide
Summary of Contents for Quattro Ultima
Page 2: ...Quattro Ultima User s Guide...
Page 68: ......
Page 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Page 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Page 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Page 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Page 222: ...Reference Information Page 222 Quattro Ultima User s Guide...