Nitrogen
A supply of dry, oil-free nitrogen at 6-7 bar (90-100 psi) is required.
Caution:
The lines supplying nitrogen to the instrument must be clean and dry.
If plastic tubing is used it must be made of Teflon. The use of other types of
plastic leads to contamination of the instrument.
CID Gas
Argon is required as collision gas. The supplied gas should be dry, of high purity
(99.9%) and at a pressure of approximately 350 mbar (5 psi).
Caution:
Operating with the CID gas at a significantly higher pressure results in
a fault.
Hardware Specifications
Page 14
Quattro Ultima
User's Guide