Check that a stable beam of solvent ions is now apparent.
Refer to
Hints for Sample Analysis
later in this chapter for further information
on source tuning.
Warning:
It is normal for the source enclosure, the glass tube and parts of the
probe adjustment flange to reach temperatures of up to 60°C during prolonged
APcI operation. Care should be exercised when handling source components
immediately after operation.
Warning:
Switch off the liquid flow and allow the probe to cool (<100°C)
before removing it from the source.
Caution:
Failure to employ a desolvation gas flow during APcI operation may
lead to heat damage to the source.
Calibration
Having obtained a stable APcI beam, refer to
Mass Calibration
earlier in this manual.
Atmospheric Pressure Chemical Ionisation
Page 173
Quattro Ultima
User's Guide