Starting the Calibration Process
To start the calibration process:
Select
OK
from the Automatic Calibration dialog box.
The instrument acquires all of the calibration files in the following order using the
data file names shown:
MS1 static calibration
data file: STATMS1
MS1 scanning calibration
data file: SCNMS1
MS1 scan speed compensation
data file: FASTMS1
MS2 static calibration
data file: STATMS2
MS2 scanning calibration
data file: SCNMS2
MS2 scan speed compensation
data file: FASTMS2
Once all of the data have been acquired each data file is combined to give a single
spectrum which is then compared against the reference spectrum to form a calibration.
This process takes place in the same order as above. If the full calibration dialog box
is open then a constantly updated status message for the calibration is displayed.
If, when the process is completed, the calibration statistics meet with the requirements
specified by the selected calibration parameters then a successful calibration message
is displayed. A calibration report is then printed showing a calibration curve for each
of the calibration processes. An example of a calibration report is shown on the
following page.
Mass Calibration
Page 114
Quattro Ultima
User's Guide
Summary of Contents for Quattro Ultima
Page 2: ...Quattro Ultima User s Guide...
Page 68: ......
Page 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Page 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Page 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Page 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Page 222: ...Reference Information Page 222 Quattro Ultima User s Guide...