Removing and Cleaning the Ion Block
On the tune page select
Other
from the menu bar at the top of the tune page.
Click on
Vent
.
The rotary pump and the turbomolecular pumps switch off. The turbomolecular
pumps are allowed to run down to 50% speed after which a vent valve opens to
atmosphere automatically.
Remove the source enclosure, sample cone and cone gas nozzle as described in
the previous section.
Maintenance and Fault Finding
Page 188
Quattro Ultima
User's Guide
Ion
Block
Cover
Plate
O Ring