Obtaining a Beam
Ensure that the corona discharge pin is fitted and connected as described in
Routine Procedures, Preparation for APcI Operation.
Ensure that the APcI probe is fitted as described above, that the desolvation gas
tube is connected to the front panel, and that the purge gas outlet is plugged.
If necessary, change the ionisation mode using the
Ion Mode
command.
The top line of the tune page indicates the current ionisation mode.
Set
Source Temp
to 130°C.
Set
APcI Probe Temp
to 20°C with no liquid flow and the nitrogen off.
Set
Corona
to 2µA and
Cone
to 50V.
When
Source Temp
reaches 130°C:
Select
Gas
,
Gas
to switch on the nitrogen gas.
Using the valves on the front of the instrument, adjust
Desolvation Gas
to
150 litres/hour and set
Nebuliser Gas
to its maximum setting.
To monitor the flow rate, select
the
Source
window on the tune page and
observe the
Gas Flows
readback window.
Select one of the peak display boxes and set
Mass
to 50 and
Span
to 90.
Select
.
Increase
Gain
on the peak display
box in the range 1 to 20 until peaks
become clearly visible.
Set
APcI Probe Temp
to 500°C.
When
APcI Probe Temp
reaches
500°C:
Start the LC pump at a flow of
1 ml/min.
Adjust the probe’s in / out position
so that it is fully retracted.
Adjust the probe’s sideways
position so that the spray is directed
approximately at the midpoint between the corona pin and the sample cone.
Atmospheric Pressure Chemical Ionisation
Page 172
Quattro Ultima
User's Guide
In / Out
Probe
Adjustment
Sideways
Probe
Adjustment
Summary of Contents for Quattro Ultima
Page 2: ...Quattro Ultima User s Guide...
Page 68: ......
Page 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Page 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Page 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Page 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Page 222: ...Reference Information Page 222 Quattro Ultima User s Guide...