Preventive Maintenance Check List
6
Avoid venting the instrument when the rotary pump is gas ballasting.
6
Do not gas ballast the rotary pump for more than 2 hours under any
circumstances.
For full details of the following procedures, consult the relevant sections of this
chapter and/or refer to the manufacturer’s literature.
Weekly
•
Gas ballast for at least 30 minutes by rotating the gas ballast knob
anticlockwise by 5 to 6 turns.
When gas ballast is complete, check the rotary pump oil level and colour.
Oil that has become noticeably red in colour should be replaced.
•
Check the water chiller level and temperature (if fitted).
Monthly
•
Check all cooling fans and filters.
•
Change the odour element in the oil mist filter.
Three-Monthly
•
Change the sorbent in the foreline trap.
Four-Monthly
•
Change the mist element in the oil mist filter.
•
Change the oil in the rotary pump.
Gas ballast lightly for 30 to 60 minutes both before and after changing oil.
Maintenance and Fault Finding
Page 211
Quattro Ultima
User's Guide