Rear Panel Connections
Event Out
Four outputs,
Out 1
to
Out 4
, are provided to allow various peripherals to be
connected to the instrument. Switches
S1
to
S4
allow each output to be set to be
either a contact closure (upper position) or a voltage output (lower position).
Out 1
and
Out 2
, when set to voltage output, each have an output of 5 volts. The
voltage output of both
Out 3
and
Out 4
is 24 volts.
During a sample run
Out 1
closes between acquisitions, and is used typically to
enable an external device to inject the next sample. The three remaining outputs are
reserved for future developments.
Contact Closure In
In 1
and
In 2
inputs are provided to allow external device to start sample acquisition
once the device has performed its function (typically sample injection).
Analog Channels
Four analog channel inputs are available, for acquiring simultaneous data such as a
UV detector output. The input differential voltage must not exceed one volt, though
full scale automatically adjusts from 1mV to 1V.
Instrument Description
Page 26
Quattro Ultima
User's Guide
REFER TO
MANUAL BEFORE
CONNECTING TO
THESE PORTS
CONTACT
CLOSURE
IN
SCOPE
MUX
DATA SYSTEM
X
Y
ANALOG CHANNELS
EVENT OUT
S4
S3
S2
S1
OUT 4
OUT 3
OUT 2
OUT 1
IN 2
IN 1
CH1
!
1V
CH2
1V
+ -
+ -
CH4
1V
CH3
1V
CAUTION !
Summary of Contents for Quattro Ultima
Page 2: ...Quattro Ultima User s Guide...
Page 68: ......
Page 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Page 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Page 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Page 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Page 222: ...Reference Information Page 222 Quattro Ultima User s Guide...