Verification
Once a full instrument calibration is in place
it is not always necessary to repeat the full
calibration procedure when the instrument is
next used. Instead a calibration verification
can be performed. (There is no benefit in
verifying each calibration individually,
re-calibration is just as quick.)
If a scanning acquisition is to be made and
the calibration is to be checked:
Set up the instrument and access the
calibrate dialog box as though a full
calibration is to be carried out.
Set all peak matching parameters to the
values that were used for the calibration.
Bring up the Automatic Calibration
dialog box by selecting
Start...
on the
Calibrate dialog box.
Select
Scanning Calibration
and deselect
Static Calibration
and
Scan Speed Compensa
tion
.
Deselect
Acquire & Calibrate
and
select
Acquire & Verify
and
Print Report.
Select either
MS1
or
MS2
,
depending on the type of
acquisition to be
performed.
Select
Acquisition Parameters
to call up the Calibration
Acquisition Set-up dialog
box, as shown.
The parameters entered should be identical to the parameters originally used for
the calibration being verified.
Mass Calibration
Page 122
Quattro Ultima
User's Guide
Summary of Contents for Quattro Ultima
Page 2: ...Quattro Ultima User s Guide...
Page 68: ......
Page 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Page 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Page 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Page 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Page 222: ...Reference Information Page 222 Quattro Ultima User s Guide...