Blockage of the stainless steel sample capillary may occur if the desolvation
heater is left on without liquid flow. This is particularly relevant for samples
contained in involatile solvents or high analyte concentrations. To avoid this
problem it is good practice to switch off the heater before stopping the liquid
flow, and flush the capillary with solvent.
A blocked stainless steel sample capillary can often be cleared by
removing it and reconnecting it in the reverse direction, thus flushing out
the blockage.
•
Gas flow problems
Check all gas connections for leaks using soap solution, or a suitable leak
searching agent such as Snoop.