Replacing the Fused Silica Capillary
With the probe removed from the source proceed as follows:
Remove the probe tip assembly and the heater, as described in the preceding
section.
Remove the probe end cover by removing the two screws and the grub screw
that retains the LC filter.
Loosen the filter from the adapter nut.
Unscrew the adapter nut from the probe.
Remove and discard the fused silica capillary.
Using a ceramic capillary cutter, cut a new length of 300µm o.d. × 100µm i.d.
fused silica capillary, about 1 centimetre excess in length.
Using a GVF/004 ferrule and the adapter nut, connect the sample capillary to the
filter ensuring that the liner tube is fully butted into the filter.
Maintenance and Fault Finding
Page 202
Quattro Ultima
User's Guide
0.5 to
1mm
Fused Silica
Capillary
Finger-tight
Nut & Ferrule
Rheodyne
Nut & Ferrule
GVF/004
Ferrule
Adapter
Nut
Filter
PTFE
Tube
Grub
Screw
Grub
Screw