Auxiliary Output Signals
In addition to a calibration signal, the following signals can be output through the AUX OUTPUT
connector:
Square Wave
Trigger Out - can be used to trigger another scope
DC level - a reference level
Trigger Enabled - can be used as a gating function to trigger another instrument when
the scope is ready
Pass/Fail - allows you to set a pulse duration from 1 ms to 500 ms; generates a pulse
when pass/fail testing is active and conditions are met.
Aux Output Off - turns off the auxiliary output signal
Auxiliary Output Setup
1. In the menu bar, touch
Utilities
, then
Utilities Setup...
in the drop-down menu.
2. Touch
the
Aux Output
tab.
3. If you simply want a 1 kHz, 1 V square wave, touch the button so labeled.
4. If you want a specialized output, touch one of the buttons under
Use Auxiliary Output For
.
5. Touch
inside
the
Amplitude
data entry field and enter a value, using the pop-up numeric
keypad. If you want a TTL level signal, touch the
TTL Level
checkbox. The
Amplitude
field
will accordingly become unavailable.
6. If you selected Square Wave, touch inside the
Frequency
data entry field and enter a value,
using the pop-up keypad. You can set a value from 5.0 Hz to 5 MHz.
7. If you selected Pass/Fail, touch inside the
Pulse Duration
field and enter a value from 1 ms
to 500 ms, using the pop-up numeric keypad.
60
SDA-OM-E Rev H
Summary of Contents for SDA
Page 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Page 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Page 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Page 246: ...246 SDA OM E Rev H ...
Page 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Page 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Page 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Page 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Page 328: ...328 SDA OM E Rev H ...
Page 394: ...394 SDA OM E Rev H ...