Interval Trigger that triggers when the interval falls within the selected range:
t1 = range's lower time limit; t2 = range's upper limit. The broken upward-pointing arrow indicates a potential
trigger, while the bold one indicates where the actual trigger occurs on the positive edge within the selected
range.
Interval Trigger Setup
1. In the menu bar, touch
Trigger
, then
Trigger Setup...
in the drop-down menu.
2. Touch
the
Interval
trigger button
.
3. Touch
inside
the
Trigger On
data entry field and select a source on which to trigger.
Touch inside the
Level
data entry field and enter a value using the pop-up numeric keypad.
4. Select positive or negative slope.
5. Touch
the
LessThan
button and enter a pulse-width value in the
Upper Limit
data entry
field.
Or
touch the
GreaterThan
button and enter a value in the
Lower Limit
data entry field.
Or
touch the
InRange
button.
88
SDA-OM-E Rev H