Transition Mode Display
Setup and Installation
Multi-Eye Setup and Installation
The Multi Eye Measurement Tool does not require any additional software installation, as it is fully
integrated into the SDA software. However, depending on the specific mode of operation
selected, each of the three new modes of operation requires distinct instrument and channel
setups.
For example, the FSB mode requires setting up multiple channels to carry out the eye diagram
measurements, as described in the example below. The Gated Eye Diagram requires at least two
inputs: The Data and the Gate signal. The Transition Mode requires at least one input (the Data
Stream, but depending on the specific serial data standard, a clock input may also be required to
recover the clock information).
These signals (except for the strobes) are differential signals, so the use of differential probes is
often required to carry out the measurements. As with any high-speed precision measurements,
all probes and SMA cables should be calibrated and deskewed prior to any data collection.
356
SDA-OM-E Rev H
Summary of Contents for SDA
Page 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Page 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Page 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Page 246: ...246 SDA OM E Rev H ...
Page 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Page 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Page 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Page 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Page 328: ...328 SDA OM E Rev H ...
Page 394: ...394 SDA OM E Rev H ...