SDA Operator’s Manual
Interval Trigger that triggers when the interval width is larger than the set interval. The broken upward-pointing
arrow indicates a potential trigger, while the bold one shows where the actual trigger occurs on the positive edge
after the selected interval.
Interval In Range:
This Interval Trigger is generated whenever an interval between two edges of
the same slope falls within a selected range. The instrument initializes and restarts the timing for
the interval whenever the selected edge occurs. You can set an interval from 2 ns to 20 s.