7. Touch
the
Graph
button, then touch inside the
Graph with
field to select a graph mode. The
Graph modes are as follows:
Histogram of the values of a parameter
Track of the values of a parameter
Trend of the values of a parameter
Resampling To Deskew
Deskew whenever you need to compensate for different lengths of cables, probes, or anything
else that might cause timing mismatches between signals. Resample a signal on one channel
and adjust it in time relative to a signal on another channel.
Resampling
1. In the menu bar, touch
Math
, then
Math Setup...
in the drop-down menu.
2. Touch a math function trace tab
F1
through
Fx
The number of math traces available depends
on the software options loaded on your scope. See Specifications..
3. Touch
the
single
function button.
4. Touch
inside
the
Source1
field and select a source: channel, math trace, memory location.
5. Touch
inside
the
Operator1
field and select
Deskew
from the
Functions
category.
6. In the dialog on the right, touch the
Deskew
tab.
7. Touch
inside
the
Delay by
data entry field and type in a time value, using the pop-up keypad.
Rescaling and Assigning Units
This feature allows you to apply a multiplication factor (
a
) and additive constant (
b
) to your
waveform:
a
X +
b
. You can do it in the unit of your choice, depending on the type of application.
Allowable unit abbreviations are as follows:
(blank) no
units
A Ampere
C Coulomb
CYCLE cycles
DB Decibel
DBC
Decibel referred to carrier
DBM Decibel
Milliwatt
DBV Decibel
Volts
DBUZ Decibel
Microamp
DEC Decade
DIV Divisions
180
SDA-OM-E Rev H
Summary of Contents for SDA
Page 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Page 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Page 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Page 246: ...246 SDA OM E Rev H ...
Page 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Page 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Page 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Page 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Page 328: ...328 SDA OM E Rev H ...
Page 394: ...394 SDA OM E Rev H ...