SDA Operator’s Manual
Hold Off by Events
Select a positive or negative slope and a number of events. An event is the number of times the
trigger condition is met after the last trigger. A trigger is generated when the condition is met after
this number, counted from the last trigger. The count is restarted on each trigger. For example, if
the event number is two, the trigger will occur on the third event. From one to 1,000,000,000
events can be selected.
Edge Trigger with Holdoff by Events (in this example, two events). The bold edges on the trigger source indicate
that a positive slope has been selected. The broken, upward-pointing arrows indicate potential triggers, while the
bold ones show where triggers actually occur after the holdoff expires.
Simple Triggers
Edge Trigger on Simple Signals
The instrument uses many waveform capture techniques that trigger on features and conditions
that you define. These triggers fall into two major categories:
•
Edge
activated by basic waveform features or conditions such as a positive or negative
slope, and hold-off
•
SMART Trigger®
sophisticated triggers that enable you to use basic or complex conditions
for triggering.
Use Edge Triggers for simple signals, and the SMART Triggers for signals with rare features, like
glitches.
SDA-OM-E Rev H
79
Summary of Contents for SDA
Page 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Page 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Page 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Page 246: ...246 SDA OM E Rev H ...
Page 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Page 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Page 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Page 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Page 328: ...328 SDA OM E Rev H ...
Page 394: ...394 SDA OM E Rev H ...