SDA Basic Setup
Note:
Many of the measurements in the SDA require both a high sampling rate and long memory to compute accurately.
Verify that you are in two channel mode (20 GS/s) and that at least 400k samples are being captured before performing
any SDA measurements. Lower sampling rates can result in less accurate jitter measurements, and short record lengths
can give incomplete eye patterns or jitter displays that diverge.
1. To access the SDA dialogs, press the Serial Data front panel button; or, touch
Analysis
in
the menu bar, then
Serial Data
in the drop-down menu. The display defaults to the Scope
dialog, which enables you to set up SDA measurements while you continue to view currently
displayed waveforms. Touching the
Scope
button turns off SDA features. In the image
above are shown two measurement modes:
Edge-Ref
and
Edge-Edge
. The Edge-Edge
mode is optional and only available with the ASDA-J option. If this option is not present, the
two mode buttons are not displayed. The Edge-Edge mode is explained separately in this
manual; the information presented here applies mainly to the standard (Edge-Reference)
mode.
2. Touch
inside
the
Data Source
field and select a data source, then touch inside the
Clock
Source
field and select a clock input. If you want to recover the clock from the signal, touch
the
Recover clock
checkbox; the
Clock Source
field will become inactive. The "Crossing
level(s)" section in this dialog allows you to set the voltage level at which the signal timing is
measured. The crossing level is set separately for the data and clock (if an external clock is
selected) and can be either absolute or relative. The
Absolute
crossing level in volts (or
watts for an optical signal) can be set directly, or can be found automatically by touching the
Find Data Level
button. The level is found by locating the midpoint between the highest and
lowest signal levels in the current acquisition. The
Relative
level is automatically set to the
selected percentage on each acquisition.
312
SDA-OM-E Rev H
Summary of Contents for SDA
Page 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Page 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Page 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Page 246: ...246 SDA OM E Rev H ...
Page 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Page 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Page 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Page 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Page 328: ...328 SDA OM E Rev H ...
Page 394: ...394 SDA OM E Rev H ...