The inputs to Param Script can also be math (F1-Fx) or memory (M1-Mx) traces. The inputs to P
Script can be the results of any parameter measurement, not necessarily Param Script.
Parameter Math Setup
1. Touch
Measure
in the menu bar, then
Measure Setup...
in the drop-down menu.
2. Touch
the
My Measure
button in the "Measure" dialog.
3. Touch
the
Px
tab for the parameter to which you want to apply parameter math.
4. In the "Px" dialog, touch the
math on parameters
button
. The
Source
field will
expand to two fields.
5. Touch
inside
the
Source1
and
Source2
fields and select the parameters you want to apply
math to (P1 to Px). If you are applying math to a single parameter (for example, invert), just
touch inside the
Source1
field and select a parameter (P1 to Px).
6. Touch
inside
the
Math Operator
field and select a math operation from the
Select
Measurement
menu. If you select an operation that requires two input parameters, the
Source
field will expand to two fields.
Parameter Script Math Setup
1. Touch
Measure
in the menu bar, then
Measure Setup...
in the drop-down menu.
2. Touch
the
My Measure
button in the "Measure" dialog.
3. Touch
the
Px
tab for the parameter to which you want to apply parameter math.
144
SDA-OM-E Rev H
Summary of Contents for SDA
Page 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Page 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Page 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Page 246: ...246 SDA OM E Rev H ...
Page 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Page 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Page 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Page 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Page 328: ...328 SDA OM E Rev H ...
Page 394: ...394 SDA OM E Rev H ...