SDA Operator’s Manual
Qualified Parameters
Some LeCroy instruments and software packages give you the ability to constrain parameter
measurements to a vertically or horizontally limited range, or to occurrences gated by a second
waveform. Furthermore, both constraints can operate together. This capability enables you to
exclude unwanted characteristics from your measurements. It is much more restrictive than
Measure Gate, which is used only to narrow the span of the waveform for analysis, along the
horizontal axis.
Note:
Since this feature operates on only a subset of the data, possible alerts or status indicators concerning the
measurement (such as " Data range too low") are not displayed.
Range Limited Parameters
Range Qualifier Setup
1. From the menu bar, select
Measure
, then
Measure Setup...
from the drop-down menu.
2. Touch
a
Px
tab to open the setup dialog.
3. Touch
inside
the
Source
field and select a source from the pop-up menu.
4. Touch
inside
the
Measure
field and select a parameter from the pop-up menu.
5. Touch
the
Accept
tab of the right-hand dialog, then touch the
Values In Range
checkbox.
Depending on whether you select a vertical or horizontal parameter, the correct units will be
automatically displayed (V, s, Hz, dB) in the
Between
and
And
fields. Or, if you select a
simple ratio parameter (such as power factor) that yields a dimensionless number, no units
will be displayed.
6. Touch
the
Find Range
button to quickly display the most recent value of the parameter
measurement. From there it is a simple matter to set the desired range.
Amplitude Parameter
(range set manually)
Delay Parameter
(Find Range selected)
SDA-OM-E Rev H
177
Summary of Contents for SDA
Page 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Page 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Page 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Page 246: ...246 SDA OM E Rev H ...
Page 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Page 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Page 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Page 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Page 328: ...328 SDA OM E Rev H ...
Page 394: ...394 SDA OM E Rev H ...