Parameter
Description
Definition
Notes
TAA-
Average local baseline-to-trough
amplitude for all local features.
Hysteresis argument used to
discriminate levels from noise
in data.
Available with DDM2 option.
Standard in DDA-5005A.
TAA+
Average local baseline-to-peak
amplitude for all local features.
Hysteresis argument used to
discriminate levels from noise
in data.
Available with DDM2 option.
Standard in DDA-5005A.
TIE@level
Difference between the measured
times of crossing a given slope and
level and the ideal expected time.
For Slope you can choose positive,
negative, or both. For output units
you can choose time or unit interval
(UI). A unit interval equals one
clock period.
The Virtual Clock setup gives you a
choice of Standard (1.544 MHz) or
Custom reference clocks.
You can also use a mathematically
derived Golden PLL to filter low
frequency jitter. The cutoff
frequency is user selectable.
Cutoff Freq =
(1/1.667e3) x Clock
Freq
Reference levels and edge-
transition polarity can be
selected. Hysteresis argument
used to discriminate levels
from noise in data.
Available with JTA2, ENET,
SDA, and XMAP options.
Standard in SDA100G,
WavePro 7000A, WaveExpert,
and sampling scopes.
Time@edge
Measures time at each edge on
each digital line
Available
with
MS-32
option.
Time@level
Time at level: Time from trigger
(t=0) to crossing at a specified
level.
Time from trigger to
crossing level
Reference levels and edge-
transition polarity can be
selected. Hysteresis argument
used to discriminate levels
from noise in data.
Standard parameter.
Top
Higher of two most probable states,
the lower being base; it is
characteristic of rectangular
waveforms and represents the
higher most probable state
determined from the statistical
distribution of data point values in
the waveform.
Value of most
probable higher state
Gives similar result when
applied to time domain
waveform or histogram of data
of same waveform. But with
histograms, result may include
contributions from more than
one acquisition.
Standard parameter.
Total Jitter
Total jitter at a given bit error rate
Available with ENET and SDA
options.
174
SDA-OM-E Rev H
Summary of Contents for SDA
Page 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Page 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Page 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Page 246: ...246 SDA OM E Rev H ...
Page 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Page 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Page 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Page 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Page 328: ...328 SDA OM E Rev H ...
Page 394: ...394 SDA OM E Rev H ...