The TIE histogram shows the measured jitter distribution that is the source for the bathtub curve
and total jitter parameter.
1. Touch
the
Jitter Histogram
button. The histogram will be displayed in the center of the
bathtub curve.
2. Touch
the
Bathtub
button to toggle the histogram display off.
DDJ (Synchronous N-Cycle Plot)
The DDj (data dependent jitter) is measured on the data stream when the
Conventional
jitter
breakdown method is selected. The DDj is measured by finding the average crossing time for
each edge in the data pattern and comparing this to the nominal crossing time, based on the
expected bit rate of the signal under test. This measurement requires a repeating data pattern to
operate. If the data pattern in the signal does not repeat at least once in the acquired data record,
the DDj parameter will indicate zero even though this type of jitter is present. A graphical display
of the DDj can be viewed by clicking the
ISI
button in the
Jitter
menu. This display shows the
position error of the average crossing point for each edge in the pattern.
1. Touch
inside
the
DDj Calc Method
control and select
Synch N Cycle
for a repeating pattern
(available for Edge-Ref or Edge-Edge method).
338
SDA-OM-E Rev H
2. Touch
the
ISI
button.
Summary of Contents for SDA
Page 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Page 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Page 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Page 246: ...246 SDA OM E Rev H ...
Page 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Page 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Page 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Page 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Page 328: ...328 SDA OM E Rev H ...
Page 394: ...394 SDA OM E Rev H ...