3. There are two crossing-level controls in the SDA main dialog: one for the data signal under
test and another for the reference clock (if the
Recover clock
checkbox is unchecked). The
crossing time used by both the jitter and eye pattern measurements is determined as the
absolute time at which the signal level crosses the selected threshold. Touch inside the
Data
Level Is
or
Clock Level Is
field and select either
Absolute
or
Percent
from the pop-up
menu.
4. If you selected
Absolute
, touch inside the
Abs level
field and enter a voltage value.
Alternatively, you can touch the
Find Data Level
or
Find Clock Level
button. If you selected
Percent, touch inside the
Percent level
field and enter a percentage value, using the pop-up
keypad. The level is automatically set to the selected percentage on each acquisition.
Note:
The absolute level setting should be used for jitter measurements over several acquisitions because the relative
setting can potentially remove jitter by tracking slowly, thereby varying level shifts that could occur between acquisitions.
5. Touch
inside
the
Signal Type
field and select a standard signal type from the pop-up menu.
The signal type defines the compliance masks and bit rates for the selected standard. If you
selected
Custom
from the
Signal Type
menu, touch inside the
Bit Rate
field and enter a bit
rate, using the pop-up keypad. By touching the
Find Bit Rate
button, you can have the bit
rate determined automatically. This button is also available when a standard data type is
selected.
SDA-OM-E Rev H
313
Summary of Contents for SDA
Page 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Page 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Page 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Page 246: ...246 SDA OM E Rev H ...
Page 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Page 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Page 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Page 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Page 328: ...328 SDA OM E Rev H ...
Page 394: ...394 SDA OM E Rev H ...