SDA Operator’s Manual
Delay Control
As you change the timebase, this control allows you to either keep the horizontal offset indicator
stationary (when
Div
is selected) or to have it move with the trigger point (when
Time
is
selected). The advantage of selecting
Div
is that the trigger point will remain on the grid as you
increase the timebase; whereas, if
Time
is selected, the trigger point could move off the grid.
Note:
Regardless of whether you select
Time
or
Div
, the "Delay" shown in the timebase setup dialog always indicates
time. However, when
Div
is selected for Delay In, the delay in time is scaled proportional to the change in timebase,
thereby keeping the division on the grid constant.
1. In the menu bar touch
Utilities
; then touch
Preferences
in the drop-down menu.
2. Touch
the
Acquisition
tab.
3. Under
Delay Setting constant in:
, touch either the
Div
or
Volts
button.
Trigger Counter
Checking the
Reset trigger counter before starting a new acquisition
checkbox clears the
trigger counter each time the scope issues an arm acquisition command. This applies when you
have set a trigger Holdoff condition in the Trigger dialog in either time or events:
The default condition of this control is off (unchecked).
Performance Optimization
You can set up the scope to optimize either calculating speed or display speed. If the display
update rate is of primary concern to you, optimize for Display. If acquisition and analysis are more
important, optimize for analysis. Optimizing for analysis can be useful when persistence or
averaging is used, giving higher priority to waveform acquisition at the expense of display update
rate.
The choices are presented as a spectrum with highest values at the extremes:
1. In the menu bar touch
Utilities
; then touch
Preferences
in the drop-down menu.
2. Touch one of the optimization icons.
SDA-OM-E Rev H
215
Summary of Contents for SDA
Page 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Page 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Page 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Page 246: ...246 SDA OM E Rev H ...
Page 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Page 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Page 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Page 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Page 328: ...328 SDA OM E Rev H ...
Page 394: ...394 SDA OM E Rev H ...