96
KHA3000
61000-3-2
Measurement Time, Margin Comment Input and Standard Name
No.
Menu item
Description
Default
1
Measurement Time (s)
Enter a measurement time. The input range is 1 to 9600 seconds.
Enter the value with the numeric keypad and press the ENTER key,
or use the large knob to increase or decrease the value.
If you set the standard to IEC 61000-3-2 Ed5.0 Class C or JIS
C61000-3-2 2019 Class C, and set the limit values to 3rd/5th/
Current Wave, the measurement time will be 0.2 seconds.
150 seconds
2
Margin (%)
Enter a margin for a limit value. The input range is 10 to 100 (for
example, enter 80 to set 80 % of the standard limit value).
Enter the value with the numeric keypad and press the ENTER key,
or use the large knob to increase or decrease the value.
100 %
3
Input 1-char
Use this item to enter characters for Memo, Model name, Type and
Serial No. in report printouts. Up to 20 alphanumeric characters and
up to 10 hiragana and katakana characters can be input. For details
on the procedure, see page 97.
–
4
Delete 1-char
Use this item to delete characters of Memo, Model name, Type and Serial
No. from report printouts. For details on the procedure, see page 97.
–
5
Change char type
Use this item to change the character types of Memo, Model name,
Type and Serial No. Every time the F3 key is pressed, the character
types switch to alphanumeric, hiragana and katakana characters in
this order.
–
6
Return
Terminates the character input, character deletion and character
type change.
–
7
Observation Period
(Tobs)
Select the operation type of the EUT as a condition for setting an
observation period (measurement time for this product). The
selected item is printed on reports. Measurement results and
judgment are not affected.
Quasi-stationary
F1
F2
Measurement Time (s)
(Input numeric value)
F3
Margin (%)
(Input numeric value)
Comment Input
Sub Menu
F4
F1
Memo
Sub Menu
F2
Model name
Sub Menu
F3
Type
Sub Menu
F4
Serial No.
Sub Menu
F5
Observation Period (Tobs)
Quasi-Stat
Short Cyclic
Random
Long Cyclic
F6
Return
F1
Input 1-char
F2
Delete 1-char
F3
Change char type
Alphanumeric
Hiragana
Katakana
F6
Return
F6
→
→
→
→
→
→
F1
F2
F6
Input 1-char
Delete 1-char
F1
F2
F3
F6
→
Sub Menu
→
Sub Menu
12
13
9
10
11
Std. Name Setting
Sub Menu
Std. Name Select
Standard
Setting
F5
14
Reference Impedance
Use
Unuse
Limit Value Name
(Alphanumeric entry)
(Alphanumeric entry)
Meas. Tech. Name
Return
Return
1
7
2
3
4
6
5
8
Menu 2/4
15
Prev/Next
Page 2/4
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......