1
KHA3000
39
Overview
Test System Configuration
The figure shown below describes a harmonic current and voltage fluctuation test system.
The test system can be configured with the AC Power Supply and Line Impedance Network
combined with this product. The LIN3020JF or LIN1020JF line impedance network can be
combined with PCR-LE AC power supply only.
Test results are output to the screen of this product. Test conditions and test results can be
saved to a CompactFlash card (CF card) or USB flash drive. Reports can also be saved.
EUT
Data communication and control
GPIB, RS232C, USB
USB
Communication
interface
N
L3
L2
L1
G
N
W
V
U
G
N
W
V
U
G
CF card
Test results
Report
printing
Hard copy
of screen
AC power supply control
RS232C (9600 bps)
AC power supply
USB flash drive
PC
Printer
Memory slot
Remote control
KHA3000
Line impedance network
There are restrictions on the use of the following AC power sources.
•
PCR-WE
This product cannot directly control PCR-WE. To use PCR-WE, the separately sold
application software SD006-KHA Harmonics Analyzing Suite is required. PCR-WE cannot
be combined with a standard line impedance network. For the line impedance network for
PCR-WE, contact your Kikusui distributor or agent.
•
PCR-LA
PCR-LA with firmware version 3.32 or 3.33 cannot be used.
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......