13
KHA3000
319
Rem
ote Control
Test Execution (Trigger Function)
The test is executed by the following three functions:
•
Sequence 1 (ACQuire): Voltage, current and harmonic current measurement
•
Sequence 2 (RUSH): In-rush current measurement
•
Sequence 3 (TEST): Execution of the harmonic or voltage fluctuation test
Sequence 1 (ACQuire)
INIT
INIT:NAME
Starts the trigger function. The command is invalid during the analysis state.
After the INIT command is sent, measurement begins immediately when TRIG:SOUR:IMM is
specified. When TRIG:SOUR:BUS is specified, measurement begins upon a software trigger.
The INIT command abandons and nullifies the measurement data collected previously. If an
FETC? query is sent immediately after the INIT command is sent, measurement data is
returned after completion of measurement.
Command (Sequence 1)
Command
INITiate[:IMMediate][:SEQuence[1]]
INITiate[:IMMediate]:NAME {ACQuire}
Parameter Value:
ACQuire
INIT:CONT
INIT:CONT:NAME
Sets the automatic continuation mode (ON/OFF) of sequence operation. The command is
invalid during the analysis state.
●
When the automatic continuation mode is ON
When TRIG:SOUR:IMM is specified, measurement begins immediately. After the
measurement is finished, new measurement begins automatically.
When TRIG:SOUR:BUS is specified, measurement begins upon a software trigger. After
the measurement is finished, the system waits for another trigger.
●
When the automatic continuation mode is OFF
The measurement currently in progress is continued unless ABOR is sent. New
measurement is not automatically continued.
Command
INITiate:CONTinuous[:SEQuence[1]] {ON|OFF|1|0}
INITiate:CONTinuous[:SEQuence[1]]?
INITiate:CONTinuous:NAME {ACQuire},{ON|OFF|1|0}
INITiate:CONTinuous:NAME?{ACQuire}
Parameter Value:
ON(1)
Automatic continuation mode ON
OFF(0)
Automatic continuation mode OFF (default)
Response
Returns the automatic continuation mode (ON/OFF) of sequence operation in the <NR1>
format.
** RST
RST
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......