6
KHA3000
131
61000-3-12 2011: Test Flow
Executing Harm
onic Tests
Test execution
Apply the Table 4 or Table 5.
Apply the limit value table
(Table 4) of balanced
three-phase equipment with
specified conditions a, b or c.
Apply the limit value table
(Table 5) of balanced
three-phase equipment with
specified conditions d, e or f.
Apply the limit value table
(Table 3) of balanced
three-phase equipment.
Apply the limit value table
(Table 2) of equipment other
than balanced three-phase.
Rsce estimation
Re-estimation of Rsce
Test execution
AC power
supply check
and record
Is the
limit value
satisfied?
No
Yes
No
Yes
Minimum
Rsce value that
satisfies the
limit value
More than 33
33
Tables 2 or 3
Tables 4 or 5
Comparison of
Rsce values
Applied
limit value
table
Requirements of power supply
Declare Ssc that corresponds
to the minimum Rsce value
that satisfies the limit value.
Conforms to 61000-3-12 2011
when short-circuit power upon connection
with the commercial power is equal to
or greater than the declared Ssc.
Conforms to
61000-3-12 2011.
The expression of each item is simplified. For details, refer to the standards.
Set the impedance of
the power supply to at
least the estimated Rsce.
Set the impedance of the
power supply to at least
1.6 times the estimated
Rsce.
Is the minimum Rsce value that
satisfiesthe limit value equal to or
less than the estimated Rsce?
Summary of Contents for KHA3000
Page 20: ...20 KHA3000...
Page 42: ...42 KHA3000 Options...
Page 70: ...70 KHA3000 Connecting Communication Cables...
Page 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Page 120: ...120 KHA3000 Controlling the AC Power Supply...
Page 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Page 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Page 352: ...352 KHA3000 Cleaning and Checking...
Page 353: ...Specifications This chapter explains the specifications of this product 15...
Page 390: ...390 KHA3000...
Page 391: ......